Improvement of the Accuracy by the Measurement of the Electrical Cell Membrane Parameters


V. ROHLÍČEK, F. RECH

Institute of Physiology, Czech Academy of Sciences, Prague, Czech Republic



Received April 19, 201
Accepted June 29, 2001


Summary
The electrical parameters of the cell membrane are mostly estimated employing ac methods. The measurement is based on the analysis of the current(s) flowing through an access resistance and the membrane. A current/potential transducer is used at the input of the device. The parameters of this transducer, especially its feedback capacity, degrades the accuracy of the measurement and hence diminishes the suppression of mutual influences of the individual parameters. The paper suggests a possible software correction and is supplemented by remarks for practical application.


Key words
Patch clamp technique · Electrical parameters of the cell membrane · Membrane capacity · Membrane conductance · Access resistance · Dual frequency method · Exocytosis